Safety of inhaled corticosteroids delivered by plastic and metal spacers

  • I. Amirav
  • , Y. Mansour
  • , T. Tiosano
  • , S. Chamny
  • , S. Chirurg
  • , S. Oren
  • , Z. Grossman
  • , L. Kahana
  • , E. Kahan
  • , M. T. Newhouse

פרסום מחקרי: פרסום בכתב עתמאמרביקורת עמיתים

6 ציטוטים ‏(Scopus)

תקציר

Background: Because of its non-electrostatic properties the metal Nebuchamber (NC) valved holding chamber (VHC) delivers a greater mass of aerosol to the mouth than the polypropylene Aerochamber (AC) VHC. Delivery of more aerosol to the lungs may also increase systemic absorption of inhaled corticosteroids (ICS) and hypothalamo-pituitary-adrenal (HPA) suppression. Methods: Thirty children (mean 4.3 (SD 0.3) years) received 200 μg budesonide twice daily by NC or AC, both with the mask provided, in a randomised, two month crossover trial. Twenty four hour urinary free cortisol (UFC) was determined as a measure of HPA suppression. Results: UFC decreased from 42.3 (7.8) nmol UFC/nmol creatinine control to 26.2 (2.4) (p = 0.06 v control) after AC, and to 24.5 (2.5) (p = 0.04 v control) after NC (p = 0.4 AC v NC). Conclusions: Despite a greater total dose delivered to the mouth, NC is not associated with greater HPA suppression when using 400 μg/day budesonide under real life conditions in young children.

שפה מקוריתאנגלית
עמודים (מ-עד)527-528
מספר עמודים2
כתב עתArchives of Disease in Childhood
כרך88
מספר גיליון6
מזהי עצם דיגיטלי (DOIs)
סטטוס פרסוםפורסם - 1 יוני 2003
פורסם באופן חיצוניכן

טביעת אצבע

להלן מוצגים תחומי המחקר של הפרסום 'Safety of inhaled corticosteroids delivered by plastic and metal spacers'. יחד הם יוצרים טביעת אצבע ייחודית.

פורמט ציטוט ביבליוגרפי