Carbamazepine interference with an immune assay for tricyclic antidepressants in plasma

  • Dhurbassa S. Chattergoon
  • , Zul Verjee
  • , Michael Anderson
  • , David Johnson
  • , Michael A. McGuigan
  • , Gideon Koren
  • , Shinya Ito

פרסום מחקרי: פרסום בכתב עתמאמרביקורת עמיתים

39 ציטוטים ‏(Scopus)

תקציר

Background: Drug toxicological screening is commonly used as a diagnostic tool in patients with suspected toxic ingestion. False positive results due to cross-reactive compounds in drug assays may lead to misdiagnosis and mismanagement, especially when child abuse is suspected. Case Report: Two of our patients with history of ingestion of carbamazepine were tested positive on screening with the tricyclic antidepressant immunoassay. The immunoassay's known cross-reactivity for carbamazepine is reportedly as low as 0.3%. Plasma samples of our patients were initially considered positive for tricyclic antidepressants because the cross-reaction of carbamazepine gave tricyclic antidepressant concentrations as imipramine equivalent sufficiently above the assay cut-off point (20 ng/mL). Later, confirmatory urine testing of both patients using high-performance liquid chromatography was negative for tricyclic antidepressants. Conclusion: This interference has significant clinical implications, and can be avoided on urine testing using a specific chromatographic assay such as high-performance liquid chromatography.

שפה מקוריתאנגלית
עמודים (מ-עד)109-113
מספר עמודים5
כתב עתJournal of Toxicology - Clinical Toxicology
כרך36
מספר גיליון1-2
מזהי עצם דיגיטלי (DOIs)
סטטוס פרסוםפורסם - 1998
פורסם באופן חיצוניכן

טביעת אצבע

להלן מוצגים תחומי המחקר של הפרסום 'Carbamazepine interference with an immune assay for tricyclic antidepressants in plasma'. יחד הם יוצרים טביעת אצבע ייחודית.

פורמט ציטוט ביבליוגרפי