דילוג לניווט ראשי דילוג לחיפוש דילוג לתוכן הראשי

A reagentless amperometric formaldehyde-selective chemosensor based on platinized gold electrodes

פרסום מחקרי: פרסום בכתב עתמאמרביקורת עמיתים

8 ציטוטים ‏(Scopus)

תקציר

Fabrication and characterization of a new amperometric chemosensor for accurate formaldehyde analysis based on platinized gold electrodes is described. The platinization process was performed electrochemically on the surface of 4 mm gold planar electrodes by both electrolysis and cyclic voltamperometry. The produced electrodes were characterized using scanning electron microscopy and X-ray spectral analysis. Using a low working potential (0.0 V vs. Ag/AgCl) enabled an essential increase in the chemosensor's selectivity for the target analyte. The sensitivity of the best chemosensor prototype to formaldehyde is uniquely high (28180 A·M-1·m2) with a detection limit of 0.05 mM. The chemosensor remained stable over a one-year storage period. The formaldehye-selective chemosensor was tested on samples of commercial preparations. A high correlation was demonstrated between the results obtained by the proposed chemosensor, chemical and enzymatic methods (R = 0.998). The developed formaldehyde-selective amperometric chemosensor is very promising for use in industry and research, as well as for environmental control.

שפה מקוריתאנגלית
מספר המאמר503
כתב עתMaterials
כרך10
מספר גיליון5
מזהי עצם דיגיטלי (DOIs)
סטטוס פרסוםפורסם - 2017

טביעת אצבע

להלן מוצגים תחומי המחקר של הפרסום 'A reagentless amperometric formaldehyde-selective chemosensor based on platinized gold electrodes'. יחד הם יוצרים טביעת אצבע ייחודית.

פורמט ציטוט ביבליוגרפי