TY - JOUR
T1 - A novel emboli protection cannula during cardiac surgery
T2 - In vitro results
AU - Shani, Liran
AU - Cohen, Oved
AU - Beckerman, Ziv
AU - Avrahami, Idit
AU - Nir, Rony Reuven
AU - Bolotin, Gil
PY - 2014/8
Y1 - 2014/8
N2 - Objective Intraoperative cerebral events are mainly caused by emboli generated by operative manipulation of the aorta. This study aimed to delineate the distribution profiles of emboli with 2 widely used cannulae and a third novel research cannula that simultaneously produces forward flow and backward suction to extract emboli from the distal aorta during cardiac surgery. Methods The current in vitro study used a silicone model of the aortic arch and branches. The main outcome measure was the distribution profile of embolic particles of different sizes to the aortic branches; 2 commercial cannulae and a third novel cannula with and without suction were used. The research cannula was examined at different suction levels and the amount of particles retrieved was measured. Results For the research curved-tip cannula, most of the small emboli were released into the brachiocephalic trunk in the model (P <.05). For the straight-tip cannula, most of the small emboli were released into the descending aorta (P <.05). Regarding the commercial curved-tipped cannula, most of the small emboli were released into the brachiocephalic trunk (47.14% ± 4.78%; P <.05) and the medium and large emboli were predominantly released into the descending aorta. Using suction, the research cannula retrieved most of the emboli released into the aorta for all particle sizes (50%-83%; P <.05). Conclusions A straight-tip cannula may be safer in terms of cerebral embolic consequences during cardiac surgery. Furthermore, the use of the research aortic cannula may be beneficial in the cardiac surgery setting by reducing the postoperative risk for stroke.
AB - Objective Intraoperative cerebral events are mainly caused by emboli generated by operative manipulation of the aorta. This study aimed to delineate the distribution profiles of emboli with 2 widely used cannulae and a third novel research cannula that simultaneously produces forward flow and backward suction to extract emboli from the distal aorta during cardiac surgery. Methods The current in vitro study used a silicone model of the aortic arch and branches. The main outcome measure was the distribution profile of embolic particles of different sizes to the aortic branches; 2 commercial cannulae and a third novel cannula with and without suction were used. The research cannula was examined at different suction levels and the amount of particles retrieved was measured. Results For the research curved-tip cannula, most of the small emboli were released into the brachiocephalic trunk in the model (P <.05). For the straight-tip cannula, most of the small emboli were released into the descending aorta (P <.05). Regarding the commercial curved-tipped cannula, most of the small emboli were released into the brachiocephalic trunk (47.14% ± 4.78%; P <.05) and the medium and large emboli were predominantly released into the descending aorta. Using suction, the research cannula retrieved most of the emboli released into the aorta for all particle sizes (50%-83%; P <.05). Conclusions A straight-tip cannula may be safer in terms of cerebral embolic consequences during cardiac surgery. Furthermore, the use of the research aortic cannula may be beneficial in the cardiac surgery setting by reducing the postoperative risk for stroke.
UR - http://www.scopus.com/inward/record.url?scp=84904650505&partnerID=8YFLogxK
U2 - 10.1016/j.jtcvs.2014.01.001
DO - 10.1016/j.jtcvs.2014.01.001
M3 - ???researchoutput.researchoutputtypes.contributiontojournal.article???
C2 - 24521961
AN - SCOPUS:84904650505
SN - 0022-5223
VL - 148
SP - 668
EP - 675
JO - Journal of Thoracic and Cardiovascular Surgery
JF - Journal of Thoracic and Cardiovascular Surgery
IS - 2
ER -