Quantification of the hemodynamics inside a novel Synchronized Therapeutic Cardiac Assist Device for Chronic Heart Failure

نتاج البحث: فصل من :كتاب / تقرير / مؤتمرمنشور من مؤتمرمراجعة النظراء

1 اقتباس (Scopus)

ملخص

A physiological cardiac assist device (PCAD) utilizing a single cannula, that works in synchrony with the heartbeat, was developed to assist the left ventricle (LV) in chronic heart failure. The cannula is inserted through the apex of the beating LV, using a specially designed device in less than a minute. The inflow from the PCAD to the LV occurs during ejection, thereby augmenting stroke volume and stroke work. The PCAD withdraws blood from the LV through the same cannula during diastole. The study evaluates PCAD model using a computational fluid dynamics method to quantify the PCAD's four hemodynamic aspects: blood mixing inside the chamber, pressure regime in the flow field, forces on the piston (motor) and shear stresses. Simulations were performed by numerically solving the continuity and the Navier-Stokes equations, using a finite volume solver (FLUENT). The analyses were time-dependent and three-dimensional with moving boundaries. The blood was assumed incompressible and Newtonian, with laminar flow. The imposed boundary conditions were fixed pressure of 100 mmHg at the cannula inlet/outlet (LV pressure), and the piston displaced 16ml within 0.3 sec during the systole, at heart rate of 70 bpm. Good hemodynamic performances were obtained. Low shear stress (<35Pa) with a shear stress-time product smaller than 0.35 Pa-sec (smaller by one order of magnitude than the threshold for platelets activation). Good washout properties and remnant fraction of 1% after 7 cycles were observed. The good agreement of our results with these predicted values confirms the ideal mixing assumption. Pressure gradients of less than 20mmHg (i.e., no cavitations) with no permanent stagnation regions were found. The forces on the piston were relatively low (20N). The results strongly suggest that there is low probability of hemolysis or thrombosis inside this novel physiological device.

اللغة الأصليةالإنجليزيّة
عنوان منشور المضيفITRE 2006 - 4th International Conference on Information Technology
العنوان الفرعي لمنشور المضيفResearch and Education, Proceedings
ناشرIEEE Computer Society
الصفحات116-120
عدد الصفحات5
رقم المعيار الدولي للكتب (المطبوع)1424408598, 9781424408597
المعرِّفات الرقمية للأشياء
حالة النشرنُشِر - 2006
منشور خارجيًانعم
الحدثITRE 2006 - 4th International Conference on Information Technology: Research and Education - Tel-Aviv, إسرائيل
المدة: 17 أكتوبر 200618 أكتوبر 2006

سلسلة المنشورات

الاسمITRE 2006 - 4th International Conference on Information Technology: Research and Education, Proceedings

!!Conference

!!ConferenceITRE 2006 - 4th International Conference on Information Technology: Research and Education
الدولة/الإقليمإسرائيل
المدينةTel-Aviv
المدة17/10/0618/10/06

بصمة

أدرس بدقة موضوعات البحث “Quantification of the hemodynamics inside a novel Synchronized Therapeutic Cardiac Assist Device for Chronic Heart Failure'. فهما يشكلان معًا بصمة فريدة.

قم بذكر هذا