تخطي إلى التنقل الرئيسي تخطي إلى البحث تخطي إلى المحتوى الرئيسي

Optimization of endotracheal tube cuff filling by continuous upper airway carbon dioxide monitoring

  • Shai Efrati
  • , Yuval Leonov
  • , Amir Oron
  • , Yariv Siman-Tov
  • , Michael Averbukh
  • , Alex Lavrushevich
  • , Ahuva Golik

نتاج البحث: نشر في مجلةمقالةمراجعة النظراء

25 اقتباسات (Scopus)

ملخص

Inappropriate cuff filling is responsible for various complications related to the use of an endotracheal tube (ETT). In this study, we evaluated an objective, noninvasive method for continuous assessment of leak around the ETT cuff by monitoring carbon dioxide pressure (PcO2) in the upper airway. PcO2 levels were measured by capnography simultaneously between the ETT cuff and the vocal cords, at the oropharynx, and in the nares of the nose. Cuff filling was regulated by an electronic controller to achieve the minimal pressure needed to prevent CO2 leak. Feasibility of the method was assessed in a human simulator and in a porcine model. Clinical function was evaluated in 60 patients undergoing surgery, comparing the method to the standard anesthesiologist evaluation. Linear correlations were observed between the ETT cuff pressure and PcO2 level in the human simulator (R 2 = 0.954, P < 0.0001) and in the porcine model (R2 > 0.98, P < 0.0001). Iodine leak around the ETT cuff, in the porcine model, occurred only when Pco2 levels were >2 mm Hg. In the surgery patients, the mean ETT cuff pressure determined clinically by the anesthesiologist was significantly higher than the optimal cuff pressure assessed by Pco2 (25.2 ± 3.6 versus 18.2 ± 7.8 mm Hg, respectively; P < 0.001). According to these findings, optimal ETT cuff filling pressure can be identified by monitoring PcO2 at the nares or the oropharynx.

اللغة الأصليةالإنجليزيّة
الصفحات (من إلى)1081-1088
عدد الصفحات8
دوريةAnesthesia and Analgesia
مستوى الصوت101
رقم الإصدار4
المعرِّفات الرقمية للأشياء
حالة النشرنُشِر - أكتوبر 2005
منشور خارجيًانعم

بصمة

أدرس بدقة موضوعات البحث “Optimization of endotracheal tube cuff filling by continuous upper airway carbon dioxide monitoring'. فهما يشكلان معًا بصمة فريدة.

قم بذكر هذا