Erroneous reporting of coagulase-negative staphylococci as Kocuria spp. by the Vitek 2 system

  • R. Ben-Ami
  • , S. Navon-Venezia
  • , D. Schwartz
  • , Y. Schlezinger
  • , Y. Mekuzas
  • , Y. Carmeli

نتاج البحث: نشر في مجلةمقالةمراجعة النظراء

51 اقتباسات (Scopus)

ملخص

Misidentification of coagulase-negative staphylococci (CoNS) may delay appropriate treatment. We investigated 20 clinical isolates identified as Kocuria spp. by the Vitek 2 system. All were identified as CoNS by 16S rRNA gene sequencing (18 Staphylococcus epidermidis, 1 Staphylococcus haemolyticus). Four Kocuria isolates were shown to be identical to CoNS from the same patient by pulsed-field gel electrophoresis. Isolates identified by Vitek 2 as Kocuria most likely represent misidentified CoNS, and if clinically indicated, should be investigated further by genomic methods.