Ergebnisse eines prospektiven Multizentervergleichs der 2.Kolonkapselgeneration mit der Koloskopie

  • R. Eliakim
  • , K. Yassin
  • , Y. Niv
  • , Y. Metzger
  • , J. Lachter
  • , E. Ga
  • , B. Sapoznikov
  • , F. Konikoff
  • , G. Leichtmann
  • , Z. Fireman
  • , Y. Kopelman
  • , S. N. Adler
  • , H. J. Schulz

نتاج البحث: نشر في مجلةمقالةمراجعة النظراء

2 اقتباسات (Scopus)

ملخص

Background and study aims:A second-generation capsule endoscopy system, using the PillCam Colon2, was developed to increase sensitivity for colorectal polyp detection compared with the first-generation system. The performance of this new system is reported. Patients and methods:In a five-center feasibility study, second-generation capsule endoscopy was prospectively compared with conventional colonoscopy as gold standard for the detection of colo rectal polyps and other colonic disease, in a cohort of patients scheduled for colonoscopy and having known or suspected colonic disease. Colonoscopy was independently performed within 10hours after capsule ingestion. Capsule-positive but colonoscopy-negative cases were counted as false-positive. Results:104patients (mean age 49.8years) were enrolled; data from 98were analyzed. Patient rate for polyps of any size was 44%, 53% of these patients having adenomas. No adverse events related to either procedure were reported. The capsule sensitivity for the detection of patients with polyps 6mm was 89% (95% confidence interval CI 7097) and for those with polyps 10mm it was 88% (95% CI 5698), with speci ficities of 76% (95% CI 7278) and 89% (95% CI 8690), respectively. Both polyps missed by colonoscopy and mismatch in polyp size by study definition lowered specificity. Overall colon cleanliness for capsule endoscopy was adequate in 78% of patients (95% CI 6886). Conclusions:The new second-generation colon capsule endoscopy is a safe and effective method for visualizing the colon and detecting colonic lesions. Sensitivity and specificity for detecting colorectal polyps appear to be very good, sug gest ing a potential for improved accuracy compared with the first-generation system. Further pro spec tive and comparative studies are needed.

العنوان المترجم للمساهمةProspective multicenter performance evaluation of the second-generation colon capsule compared with colonoscopy
اللغة الأصليةألمانيّة
الصفحات (من إلى)144-149
عدد الصفحات6
دوريةEndoskopie Heute
مستوى الصوت23
رقم الإصدار2
المعرِّفات الرقمية للأشياء
حالة النشرنُشِر - 2010
منشور خارجيًانعم

بصمة

أدرس بدقة موضوعات البحث “Ergebnisse eines prospektiven Multizentervergleichs der 2.Kolonkapselgeneration mit der Koloskopie'. فهما يشكلان معًا بصمة فريدة.

قم بذكر هذا