A new hazard in the use of an external demand pacemaker

Y. Hasin, Y. Mahler, S. Rogel

نتاج البحث: نشر في مجلةمقالةمراجعة النظراء

1 اقتباس (Scopus)

ملخص

Ventricular pacing by a Medtronic model 5880A external demand pacemaker was shown to be inhibited as long as the metal coverplate of the pacemaker was touched with bare hands. The asystolic periods were induced by a previously undescribed malfunction of the pacemaker, inherent in its structure. This phenomenon occurred when both patient and physician were isolated from any main operated instrument. It could be demonstrated that pacer inhibition was induced by capacitive coupled 50 Hz fields from the environment, activating the sensing mechanism through the metal coverplate. The plastic covered Medtronic model 5840 pacemaker does not have such undesirable characteristics.

اللغة الأصليةالإنجليزيّة
الصفحات (من إلى)93-96
عدد الصفحات4
دوريةJournal of Electrocardiology
مستوى الصوت11
رقم الإصدار1
المعرِّفات الرقمية للأشياء
حالة النشرنُشِر - 1978

بصمة